I. Gas Path Leak Check
All Categories >
Inspection and Elimination of Air Circuit Leaks
2020-08-07
According to the strictness of air circuit sealing, the methods for checking whether there are air leaks are classified into three levels: A, B, and C.
Class A Leak Test: The most rudimentary inspection for severe air leaks. Typically, after the air source has been turned on and stabilized, you should not hear any hissing sounds indicating air leakage at the connections of various pipes and valve components along the air circuit. If you do hear a distinct hissing sound, it indicates a serious leak in the system! You must trace the source of the leak based on the sound and eliminate it immediately. Common causes of major system leaks include loose air circuit connectors, cracks in the tubing within the air circuit, and failure to use appropriate gaskets. To detect severe air leaks, you can also gradually apply soapy water to each connection while maximizing the flow rate through the circuit and observe carefully for the formation of bubbles to confirm the presence of a leak.
Class B Leak Test: This test is used to detect minor leaks in the gas circuit. The procedure involves blocking the gas outlet and observing the rotor inside the flowmeter in the gas circuit. If the rotor gradually drops to zero, the gas circuit passes the Class B leak test. If the rotor fails to reach zero, carefully apply soapy water to all connections and inspect thoroughly until the leak is located.
Class C Leak Test: This test is used to detect even the tiniest leaks in the air circuit. The procedure involves blocking the air circuit outlet and monitoring the system pressure gauge; within half an hour, the pressure should not drop by more than 5 kPa (equivalent to 0.05 kgf/cm²). At this point, the system pressure should remain above 0.25 MPa (equivalent to 2.5 kgf/cm²). If necessary, a Class 0.5 standard pressure gauge can be externally connected to the system outlet to accurately measure pressure changes.
When a leak in the air circuit system is confirmed, the scope of the leak can be narrowed down by sequentially blocking or shutting off sections of the air circuit.
The vast majority of air leaks occur at the connections in the air lines, and leaks inside the air line valves also happen from time to time. As for leaks along the pipe runs themselves, they are rare except at sharp bends.
II. Troubleshooting Airline Connector Leaks
When a leak is detected at a gas line fitting, simply tightening the fitting’s screw and nut more firmly at the leaking location may seem like a straightforward solution. While this approach can sometimes yield temporary results, in most cases it may cause permanent damage to the fitting—such as stripped threads, scratches on the sealing surface, or even fitting breakage—leading to even greater problems. The correct procedure, therefore, is to first perform the following inspection on the fitting in question whenever a leak is discovered:
(1) Are the gasket and mating surfaces properly matched, and are they annealed and free of scratches?
(2) Is the joint sealing area clean, smooth, and free of contaminants?
(3) When assembling the joints, are they properly aligned and accurately matched to each other?
(4) Can you first loosely tighten the joint by hand?
If the above-mentioned inspection reveals no abnormalities, use two wrenches to tighten the fittings. When tightening, pay attention to applying appropriate pressure. For fittings equipped with plastic, rubber, or PTFE gaskets, avoid applying excessive pressure—generally, tightening slightly beyond the point where a seal is achieved is sufficient. For fittings with metal gaskets, you may increase the tightening pressure somewhat, but still ensure that there are no air leaks.